# Device History Record (DHR) Review and Lot Release Checklist

> Source: https://netray.co/tools/device-history-record-review-and-lot-release-checklist
> Publisher: Netray (netray.co), AI and integration engineering for ERP and manufacturing systems.
> Section: Free Tools

## Short answer

A pre-release checklist for the person who signs off a device lot: verifies identification, quantities, materials, process records, acceptance activities, sterilization, labeling and open deviations.

## Category

quality

## Industry

Medical Devices

## Audience

- Quality assurance
- Batch record reviewer
- Production supervisor
- Regulatory affairs

## Tool

**Kind:** checklist

**Groups:**

#### Identification and quantities

**Items:**

**Text:** Device name, part number, revision and lot number match the work order and the label

**Critical:** true

Dates of manufacture are recorded and consistent across all pages

**Text:** Quantity started, quantity manufactured and quantity to be released are recorded and reconcile

**Critical:** true

**Text:** UDI production identifier (lot, serial, expiration, manufacture date) matches the label and records

**Critical:** true

Device master record or work instruction revision used was the effective revision

#### Materials and traceability

**Items:**

**Text:** Every component and material lot is recorded and each was released before use

**Critical:** true

Component lots are within expiration and shelf life at time of use

**Text:** Traceability to critical components, equipment and operators is complete (7.5.9)

**Critical:** true

Material issue and return quantities reconcile with the bill of materials

Line clearance and changeover records exist to prevent mix-ups

#### Process records

**Items:**

**Text:** Every step is initialed and dated by a trained operator, with no blank fields

**Critical:** true

**Text:** Process parameters were within validated ranges and monitoring records are attached

**Critical:** true

Equipment used is identified and calibration was in date on the day of use

Environmental monitoring results for controlled areas are within limits and reviewed

Corrections are single-line, initialed, dated and explained, with no overwriting or white-out

#### Inspection, test and acceptance

**Items:**

**Text:** In-process and final inspection records show sample plan, results and inspector

**Critical:** true

All results meet acceptance criteria, and measured values are recorded as required

Test equipment and gauges are identified and in calibration

Any retest has a documented reason and is authorized

**Text:** Acceptance records show who accepted the lot and when (7.5.1)

**Critical:** true

#### Sterilization, packaging and labeling

**Items:**

**Text:** Sterilization load record is attached, with cycle parameters and biological indicator results meeting criteria

**Critical:** true

Parametric or biological release is documented per the validated release method

Packaging seal integrity checks are recorded and within specification

Label reconciliation is complete: issued, used, destroyed and returned quantities

**Text:** Labeling and instructions for use revisions match the approved device master record

**Critical:** true

#### Deviations and final disposition

**Items:**

**Text:** All nonconformances raised during the lot are closed or have approved dispositions

**Critical:** true

Deviations and planned departures are approved and assessed for product impact

No open CAPA or hold applies to the materials or equipment used

Reviewer is independent of the production steps performed on the record

**Text:** Release is signed, dated and the quantity released is recorded

**Critical:** true

**Scoring Note:** Any critical item that is not satisfied stops the release until resolved. Use the checklist as a review aid, not a replacement for your device history record procedure. FDA's QMSR incorporates ISO 13485:2016 by reference, so record content is also defined by clauses 4.2.5 and 7.5.1.

## Content

**Intro:** The device history record is where a lot proves it was built as specified. The reviewer who signs the release is the last check before product leaves the building, and errors in records are among the most common findings in inspections. This checklist keeps the review consistent and focused on the items that can actually stop a release.

**Sections:**

#### How to use the checklist

**Body:** Work through the groups in order for each lot. Start with identification and quantities because errors there can invalidate the whole record. Move through materials, process, inspection and sterilization, and finish with open deviations and the release signature. Mark each item as satisfied, not applicable, or an issue, and write the issue on a defect log so it can be corrected by the original author with the proper correction method rather than by the reviewer.

**Bullets:**

- Review the record on paper or screen in one sitting where possible
- Never correct entries for the operator; return them with a note
- Use a second reviewer for implantable or high-risk devices
- Track record error types for training and trending

#### What the record must contain

**Body:** Under the former 21 CFR 820.184, the device history record covers dates of manufacture, quantity manufactured, quantity released for distribution, acceptance records showing the device was made per the device master record, the primary identification label and labeling, and any device identification and control numbers. ISO 13485 clause 7.5.1 carries the same intent, requiring a record for each device or batch that provides traceability and identifies the amount manufactured and approved for distribution. Always follow your own procedure and customer quality agreement for any additional content.

#### Using the review for improvement

**Body:** Count the issues found per record over time. A high rate of blank fields points to a form design or training problem, while recurring calibration lapses point to equipment control. Right-first-time record rate is a useful management metric, and electronic batch records can enforce completeness by making fields mandatory. Compare the cycle time of the review with your shipping schedule, since long record reviews are a common cause of late shipments, and consider review by exception for validated, stable processes.

#### Common mistakes

**Body:** Reviewers sometimes sign after only checking the final test result, missing that a step was performed out of sequence or with expired material. Another is allowing the production supervisor to review their own record. Late entries without explanation, missing initials and unreconciled labels are repeat findings. Finally, do not release against an open nonconformance on the assumption it will be closed later. The release decision must reflect the status of the lot at that time.

**Cta:** If your records are already in an ERP or MES, we can add automated completeness checks so exceptions are flagged before the reviewer opens the file.

**Faq:**

#### Who is allowed to review and release a device lot?

**Answer:** A person designated in your procedure, with authority and training for the task, who did not perform the production steps being reviewed. Many companies reserve final release for quality assurance and require a second review for sterile or implantable devices.

#### Can I release product before the DHR is fully reviewed?

**Answer:** Not under a compliant procedure. The release decision depends on evidence in the record that the lot was made and accepted as specified. Some processes use staged review during production, but the final review must precede release.

#### How should I handle a missing signature?

**Answer:** Do not let the reviewer sign for the operator. Return it to the author, who may add a late entry if they remember the activity and can justify it, or raise a nonconformance if the step cannot be verified. Document the decision.

#### Is an electronic DHR treated differently?

**Answer:** The content requirements are the same. Electronic records and signatures also need controls such as validation, audit trails, access control and, in the US, the expectations of 21 CFR Part 11 where applicable to your records.

## Search terms

DHR review checklist, batch record review checklist medical device, lot release checklist, device history record checklist, QA release review form, 21 CFR 820.184 checklist, final release review medical device, batch record error checklist

---