Measurement System and Calibration Audit Checklist

This checklist covers what auditors typically look at when they review measurement system control: identification, calibration and traceability, out-of-tolerance handling, MSA studies and records. Use it for internal audits, supplier audits or pre-audit self-checks.

0%

0 of 36 items complete

6 critical items still open - these are the highest-risk gaps.

Download checklist:

Identification and control of equipment

Calibration and traceability

Out-of-tolerance handling

MSA studies and use of data

Handling, storage and records

Critical items should all be met. Count the proportion of items met in each group. Any critical item not met is a finding. Adjust the list to match your standard and customer requirements.

Get a second pair of eyes on your gage control

Optional: an expert review of how your calibration data is stored and surfaced in your ERP.

Optional. The tool and every export work without this. No spam.

How to use it

Walk the floor with the checklist in hand. Pick a sample of gages from the master list and trace each to its label, its certificate and its last record. Then pick a few characteristics from a control plan and trace them to the gage and the MSA study. Tick items as met, and record notes for each finding. Export to PDF or Excel for the audit report.

What the standards expect

ISO 9001 clause 7.1.5 requires monitoring and measuring resources to be suitable, calibrated or verified, traceable and protected. AS9100 adds requirements on recall and impact assessment. IATF 16949 asks for MSA on all gage types in the control plan and for a calibration records set including as-received data. Customer-specific requirements often add details.

Reading the results

A pattern of missing as-found data usually means the calibration provider is not reporting it or the shop is not asking for it. A pattern of overdue gages points to a scheduling or ownership problem. A strong program has no gages in use past due, an impact process that is actually run and MSA that matches the control plan.

Common audit findings

Frequent findings include gages without IDs, certificates lacking uncertainty, intervals never adjusted, no out-of-tolerance review, and MSA done once at launch and never updated. Using this checklist before the audit gives you time to close them.

Frequently Asked Questions

Does this replace the standard?

No. It is a working aid built around common expectations in ISO 9001, AS9100 and IATF 16949. Always audit against the exact clauses, edition and customer requirements that apply to you.

How often should I audit calibration?

Many organisations include it in the annual internal audit plan, with extra checks on critical gages and a monthly review of overdue items. Risk and past findings should drive the frequency.

What is an impact review?

When a gage is found out of tolerance, you review product measured with it since the last good calibration to decide whether any of it could be nonconforming, and act on that, including customer notification if needed.

Should every instrument be calibrated?

No. Items used only for reference or indication can be excluded if the rule is documented and they are clearly labeled. Anything used to accept product or control a process should be in the program.