A design verification protocol states, before any testing, what will be tested, on how many units, by what method, and what counts as a pass. This template gives you that structure so it can be reviewed and approved ahead of the test.

Fill in

Live preview
Download:

Medical Device Design Verification Protocol Template

Design Verification Protocol

Company: Acme Medical Devices Inc. Protocol No: DVP-0042 Revision: A Device: Single-use infusion set, model IS-200 Date: 2026-10-20

1. Objective and Scope

Demonstrate that the design output meets the specified design inputs for flow rate accuracy and tubing joint tensile strength.

Design input(s) verified: DI-017, DI-018 (flow rate, tensile strength)

2. Sample Plan

Sample size: 30 units Rationale: Binomial, 95% confidence, 90% reliability, zero failures allowed

Samples must come from units built to released drawings and approved process, with lot numbers recorded on the data sheet.

3. Test Method

Gravity flow at 1 m head height with water at 23 C, collected over 10 minutes. Tensile pull at 200 mm/min on a calibrated load frame.

4. Acceptance Criteria

Flow rate 100 mL/h +/- 10%. Joint tensile strength not less than 15 N with no separation.

5. Equipment and Calibration

Load frame LF-07 (cal due 2027-02-01), balance BAL-03 (cal due 2027-01-15)

6. Results Record

Sample IDLotMeasured valuePass / FailTester initials
1
2
3

7. Deviations

Record any departure from the protocol, the reason, and the impact assessment. Overall result: Pending

Approvals

Author / Date
Design engineering / Date
Quality assurance / Date
Regulatory affairs / Date

Starting point only. Align the format with your design control procedure and keep the protocol approved before testing begins. Not legal or regulatory advice.

Want verification records tied to ERP lots?

An optional review of how to link test records to lot and equipment data.

Optional. The tool and every export work without this. No spam.

How to use the template

Fill in the fields on the left and the document updates on the right. Start with the design input reference, because every verification test should trace to a specific design input. Then set the sample size with a stated rationale, write the method in enough detail that another technician could repeat it, and state numeric acceptance criteria. Export to PDF or Word, route it for approval, and only then run the test. Results go in the table or on attached raw data sheets.

  • Trace each test to a design input ID
  • State the sample rationale, not just the number
  • List calibrated equipment by ID
  • Approve the protocol before testing

What the standards expect

ISO 13485 clause 7.3.6 and FDA 21 CFR 820.30(f) require design verification to confirm that design outputs meet design inputs, with results, the method, and the date and individuals performing it documented. FDA is moving to the Quality Management System Regulation, which incorporates ISO 13485, but the verification expectation stays the same. The protocol is the place where the plan is fixed in advance, which is what auditors look for.

Common mistakes

The most frequent finding is acceptance criteria written after seeing the data. Others include samples taken from prototype builds that differ from the released design, uncalibrated equipment, and no rationale for the sample size. Verification is also confused with validation: verification asks whether you built the device right against the inputs, validation asks whether the device meets user needs in the intended use.

  • Criteria set after testing
  • Prototype samples instead of design-frozen units
  • Missing calibration status
  • Mixing up verification and validation

Frequently Asked Questions

What is the difference between design verification and validation?

Verification confirms design outputs meet design inputs, usually through bench tests, inspection or analysis. Validation confirms the finished device meets user needs and intended use, often under actual or simulated use conditions with production-equivalent units.

How do I justify a sample size?

Use a statistical rationale tied to risk, such as a binomial confidence and reliability calculation for attribute data or a tolerance interval for variable data. Higher risk characteristics need higher confidence and reliability.

Can I change the protocol mid-test?

Only through a documented amendment or deviation with approval and an impact assessment. Silent changes are a typical audit finding.

Does this template replace my procedure?

No. It is a starting format. Follow your own design control procedure for numbering, approvals and record retention.